GENKO
Criteria for validating a GENKO application
The assessment distinguishes cycle operation, measurement quality and the relationship to the process. Each claim requires a test method and acceptance criterion for the configuration being assessed.
Operation
Check that the cycle and recording work under installation conditions. A functional test establishes that scope and date.
Measurement quality
Review representativeness, calibration, repeatability, invalid data and comparison with a reference. Performance values must be documented for each configuration.
Operational effect
Evaluating water, flocculant or stability requires a baseline, comparable conditions and a record of interventions. A favourable result is not automatically attributable to the instrument.